This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Opti4 Laparoscopic Handset

Starting date:
January 27, 2017
Posting date:
February 14, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-62200

Reason

Handset continues to operate after release of the activation button. If the device continues to operate after deactivation by the user, it can increase the potential for internal tissue burn/damage.

Affected products

Opti4 Laparoscopic Handset

Lot or serial number

More than 10 lots, contact manufacturer

Model or catalog number

E2750

Companies

Manufacturer

Covidien LLC

15 Hampshire Street

Mansfield

02048

Massachusetts

UNITED STATES