OnLine DAT Amphetamines II (2014-02-06)

Starting date:
February 6, 2014
Posting date:
March 20, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-38465

Recalled Products

OnLine DAT Amphetamines II

Reason

The manufacturer's internal measurements revealed a reduction of D-amphetamines cross-reactivity for different OnLine DAT Amphetamine II reagent lots (lots 68848302 and 68847701). The recovery of patient and control samples may be decreased with the affected lots.

Affected products

OnLine DAT Amphetamines II

Lot or serial number

04512880190

04939425190

Model or catalog number

68847701

68848301

Companies
Manufacturer

Roche Diagnostics GmbH

Sandhofer Strasse 116

Mannheim

68305

GERMANY