This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Oncor Expression Digital Linear Accelerator (2014-01-23)

Starting date:
January 23, 2014
Posting date:
March 18, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-38289

Recalled Products

Oncor Expression Digital Linear Accelerator 

Reason

The side rails on the tabletop tt-a or tt-s of the 550 TXT treatment table might have been attached incorrectly to the tabletop due by a Siemens supplier. Due to a manufacturing error, the screws holding the side rails in position might break off. As a result, the mechanical stability of the side rails cannot be guaranteed.

Affected products

Oncor Expression Digital Linear Accelerator 

Lot or serial number

5037

Model or catalog number

7360717

Companies
Manufacturer

Siemens AG

Wittelsbacherplatz 2

Muenchen

80333

GERMANY