Oncentra External Beam/Oncentral Brachy 4.5

Starting date:
December 23, 2016
Posting date:
January 12, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-61766

Reason

If a physical wedge, for which the total field size limit perpendicular to the wedge direction is more than 10 cm larger than the field size limit in wedge direction, is inserted in in-plane direction (as seen with collimator angle 0 degrees), then with very large fields perpendicular to wedge direction a part of the scattered fluence can become unmodulated by the wedge. This will result in profile "horns" close to the edge of the field.

Affected products

Oncentra External Beam/Oncentral Brachy 4.5

Lot or serial number

Version 4.5.X

Model or catalog number

170.730

Companies

Manufacturer

Nucletron B.V.

Waardgelder 1

Veenedaal

3905 TH

NETHERLANDS