ONCENTRA EXTERNAL BEAM / ONCENTRA BRACHY
Brand(s)
Last updated
Summary
Product
ONCENTRA EXTERNAL BEAM / ONCENTRA BRACHY
Issue
Medical devices - Software defect
What to do
Contact the manufacturer if you require additional information.
Affected products
Affected Products |
Lot or serial number |
Model or catalog number |
|---|---|---|
ONCENTRA EXTERNAL BEAM / ONCENTRA BRACHY |
More than 10 numbers, contact manufacturer. |
170730 |
Issue
|
Oncentra Brachy offers applicator modelling functionality, which aids the end-user in the reconstruction of an entire applicator geometry by using anchor points. In two rare scenarios, the applicator model may get mirrored.
Recall date: 2021-10-13 |
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
NUCLETRON B.V. (A WHOLLY OWNED SUBSIDIARY OF ELEKTA AB),
WAARDGELDER 1, VEENENDAAL
Netherlands, 3905 TH
Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type II
Identification number
RA-63525