Health product recall

Omni-Tract Surgical Retractor System

Last updated

Summary

Product
Omni-Tract Surgical Retractor System
Issue
Medical devices - Sterility
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare
Industry

Affected products

Affected products Lot or serial number Model or catalogue number
Omni-Tract Surgical Retractor System All lots. Not applicable

Issue

Integra Lifesciences is issuing a correction to the Integra® Omni-Tract® Surgical Retractor System - existing Instructions For Use (IFU) related to prevacuum sterilization dry time requirements, to ensure the Omni-Tract® devices and components are properly dried during sterilization to qualify current ISO standards.  The current Omni-Tract® Surgical Retractor System IFU states a dry time of 30 minutes. The validated drying times utilizing standard pre-vacuum temperature and time standards were significantly longer and needed to be extended to 90 minutes, 110 minutes, or 120 minutes depending on the sterilization cycle exposure time used.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies
Integra Lifesciences Corporation
4900 Charlemar Drive, Cincinnati, Ohio, United States, 45227
Published by
Health Canada
Audience
Healthcare
Industry
Recall class
Type II
Recall date
Identification number
RA-82277

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe