Olympus Medical Single Use Electrosurgical Snare SD-400

Last updated

Summary

Product
Olympus Medical Single Use Electrosurgical Snare SD-400
Issue
Medical devices - Sterility
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Single Use Electrosurgical Snare SD-400

More than 10 numbers, contact manufacturer.

SD-400U-15

Single Use Electrosurgical Snare SD-400

More than 10 numbers, contact manufacturer.

SD-400U-10

Issue

Olympus's investigation determined that small holes may occur during packaging of SD-400 device pouches into the shelf carton and/or during transportation of the shelf carton. Users should cease usage of the affected lots immediately due to the potential sterility breach.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies

Olympus Medical Systems Corp.

2951 Ishikawa-Cho, Hachioji-Shi, Tokyo-To, Japan, 192-8507

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-78388