Health product recall

Olympus MAJ-1443 and MAJ-1444 Valves

Last updated

Summary

Product
Olympus MAJ-1443 and MAJ-1444 Valves
Issue
Medical devices - Labelling and packaging
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Air/Water Valves

All lots.

MAJ-1444

Reusable Suction Valve

All lots.

MAJ-1443

Endoscope Reprocessor

All lots.

OER-ELITE

OER-Pro Endoscope Reprocessor

All lots.

OER-PRO

Issue

Olympus is informing that the MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OERPro and OER-Elite automated endoscope reprocessors. This means the MAJ-1443 and MAJ-1444 can no longer be high-level disinfected using the OER-Pro or the OER-Elite and must instead be reprocessed by methods described in updated IFU.
 

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Gastroenterology and urology
Health products - Medical devices - General hospital and personal use
Companies

Olympus Medical Systems Corp.

2951 Ishikawa-Cho, Hachioji-Shi, Tokyo-To, Japan, 192-8507

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-81697

Get notified