Health product recall

Olympus Endoscope Reprocessor

Last updated

Summary

Product
Olympus Endoscope Reprocessor
Issue
Medical devices - Sterility
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Endoscope Reprocessor

Contact the manufacturer.

OER-ELITE

Issue

This recall pertains to the MAJ-2318 water filter, which is compatible with the following Olympus endoscope reprocessors: OER-Elite. This water filter has been designed to prevent rinse water contamination in the OER endoscope reprocessors. Olympus confirmed there is a possibility that the water filter MAJ-2318 may have been assembled improperly. Specifically, the adapter may be installed on the incorrect end of the filter body. If customers use a defective filter, the tap water supplied to the OER endoscope reprocessors may not be properly filtered. As a result, unfiltered tap water would be used for reprocessing.

Recall start date: January 22, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Olympus Medical Systems Corp.

2951 Ishikawa-Cho, Hachioji-Shi, Tokyo-To, Japan, 192-8507

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75045