This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Olympic Pasteurmatic 3500 System (2017-05-22)

Starting date:
May 22, 2017
Posting date:
June 21, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63658

Reason

There was no initial testing completed on the Pasteurmatic products to show that the vegetative bacteria and viruses are killed in the 165 degrees fahrenheit for 30 minutes disinfection process.

Affected products

Olympic Pasteurmatic 3500 System

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

52640

52640R

52641

52641R

Companies

Manufacturer

Natus Medical Incorporated

5900 First Avenue South

Seattle

98108

Washington

UNITED STATES