This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

OHMEDA Infant Warmer System

Starting date:
August 25, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-60086

Affected Products

  1. OHMEDA Infant Warmer System
  2. Infant Warmer System

Reason

Hot screws from the "heater head" of the IWS could fall onto the bed if the "heater head" assembly has been improperly serviced. The attached service manual addendum provides instructions for checking and tightening the screws.  During each annual preventive maintenance check, continue to ensure that the screws are tight.

Affected products

A. OHMEDA Infant Warmer System

 

Lot or serial number

More than 10 lots, contact manufacturer.


 

Model or catalog number
  • 3000
  • 3050
  • 3100
  • 3150
  • 3300
  • 3400
  • 3500
  • 4000
  • 4300
  • 4400
Companies
Manufacturer

Ohmeda Medical, A Division of Datex Ohmeda Inc.

8880 GORMAN ROAD

LAUREL

Maryland

UNITED STATES



B. Infant Warmer System

Lot or serial number

More than 10 lots, contact manufacturer.

Companies
Manufacturer

Ohmeda Medical, A Division of Datex Ohmeda Inc.

8880 GORMAN ROAD

LAUREL

Maryland

UNITED STATES