OEC Elite and OEC Mobile C-arms
Brand(s)
Last updated
Summary
Product
OEC Elite and OEC Mobile C-arms
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
| Affected products | Lot or serial number | Model or catalog number |
|---|---|---|
| OEC Mobile Elite C-Arms General Surgical Platform Non-Motorized Cooling | E9-2710 E9-2720 E9-0539 | OEC 9900 ELITE |
| GE OEC Elite | FAXXTE01158 FBXXTE02995 | OEC ELITE |
Issue
GE Healthcare has become aware that recently manufactured OEC Elite and OEC Mobile C-arms, as well as recently replaced x-ray tubes on certain OEC Elite, OEC 3D and OEC 9900/9800 Mobile C-arms potentially have insufficient sealing that can result in an oil leak. This issue could result in loss of fluoroscopic imaging capability.
Recall start date: December 5, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
GE OEC Medical Systems Inc.
384 Wright Brothers Drive, Salt Lake City, Utah, United States, 84116
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76718