Health product recall

Oasis Drain, Dry Single W/AC (2015-05-07)

Starting date:
May 7, 2015
Posting date:
June 1, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-53577

Recalled Products

  1. Oasis Drain, Dry Single W/AC

Reason

In rare cases the O-ring on the in-line connector may be missing on the end of the patient line. The O-ring is a black circular ring that is clearly visible when the two pieces of the in-line connector are separated.  If this O-ring is missing, then air can pass into the patient tubing causing a system air leak.  The chest drain system will no longer be a closed system as required and you will not obtain appropriate function of the drain in evacuating air, blood or fluid.

Affected products

A. Oasis Drain, Dry Single W/AC

Lot or serial number
  • L/N 222115
  • L/N 222118
  • L/N 222528
Model or catalog number

3600-100

Companies
Manufacturer
Atrium Medical Corporation
5 Wentworth Drive
Hudson
New Hampshire
UNITED STATES