This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Oasis Chest Drain

Starting date:
January 19, 2017
Posting date:
February 9, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-62140

Reason

It has been discovered through internal testing that there have been a small quantity of drains manufactured where the valve could potentially not re-seal properly after depressing the valve. If the patient is connected to suction on an oasis drain (with a filled air-leak monitor), a manual vent valve that will not re-seal is immediately obvious if continuous bubbling in the water seal chamber is observed.

Affected products

Oasis Chest Drain

Lot or serial number

Lot # 242910

Model or catalog number

3600-100

Companies

Manufacturer

Atrium Medical Corporation

5 Wentworth Drive

Hudson

03051

New Hampshire

UNITED STATES