Oasis Chest Drain
- Starting date:
- January 19, 2017
- Posting date:
- February 9, 2017
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-62140
Reason
It has been discovered through internal testing that there have been a small quantity of drains manufactured where the valve could potentially not re-seal properly after depressing the valve. If the patient is connected to suction on an oasis drain (with a filled air-leak monitor), a manual vent valve that will not re-seal is immediately obvious if continuous bubbling in the water seal chamber is observed.
Affected products
Oasis Chest Drain
Lot or serial number
Lot # 242910
Model or catalog number
3600-100