Health product recall

Oasis Chest Drain

Starting date:
January 19, 2017
Posting date:
February 9, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-62140

Reason

It has been discovered through internal testing that there have been a small quantity of drains manufactured where the valve could potentially not re-seal properly after depressing the valve. If the patient is connected to suction on an oasis drain (with a filled air-leak monitor), a manual vent valve that will not re-seal is immediately obvious if continuous bubbling in the water seal chamber is observed.

Affected products

Oasis Chest Drain

Lot or serial number

Lot # 242910

Model or catalog number

3600-100

Companies

Manufacturer
Atrium Medical Corporation