This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

O-TWO ADULT VENTILATION TIMER (2015-06-24)

Starting date:
June 24, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-55780

Affected Products

O-TWO ADULT VENTILATION TIMER

Reason

Specific lots of the O-TWO ADULT VENTILATION TIMER may malfunction. The flash rate of the malfunctioning device may be much faster than it is labelled, with a reported flash rate of 20 per minute, instead of the labelled flash rate of 10 per minute.

Affected products

O-TWO ADULT VENTILATION TIMER

Lot or serial number
  • 078734
  • 081071
Model or catalog number

01BM1000

Companies
Manufacturer

O-Two Medical Technologies Inc.

7575 Kimbel Street

Mississauga

L5S 1C8

Ontario

CANADA