This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

O-ARM O2 IMAGING SYSTEM (2017-10-27)

Starting date:
October 27, 2017
Posting date:
November 10, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-65088

Affected products

O-ARM O2 IMAGING SYSTEM

Reason

During servicing of certain internal components of the O-arm O2, it is possible that the high-power cable connector may come in contact with surrounding metal (toggle switch or corner of battery tray) during the disconnecting/reconnecting process. If this occurs, it may result in an electrical short and a shock to the field service technician.

Affected products

O-ARM O2 IMAGING SYSTEM

Lot or serial number

All lots

Model or catalog number

BI-700-02000

Companies
Manufacturer

Medtronic Navigation, Inc. (Littleton)

300 Foster Street

Littleton

01460

Massachusetts

UNITED STATES