O-ARM O2 IMAGING SYSTEM (2017-10-27)

Starting date:
October 27, 2017
Posting date:
November 10, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-65088

Affected products

O-ARM O2 IMAGING SYSTEM

Reason

During servicing of certain internal components of the O-arm O2, it is possible that the high-power cable connector may come in contact with surrounding metal (toggle switch or corner of battery tray) during the disconnecting/reconnecting process. If this occurs, it may result in an electrical short and a shock to the field service technician.

Affected products

O-ARM O2 IMAGING SYSTEM

Lot or serial number

All lots

Model or catalog number

BI-700-02000

Companies
Manufacturer
Medtronic Navigation, Inc. (Littleton)
300 Foster Street
Littleton
01460
Massachusetts
UNITED STATES