Health product recall

NxStage System One Express Fluid Warmer (2018-06-11)

Starting date:
June 11, 2018
Posting date:
June 22, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67100

Affected products

NxStage System One Express Fluid Warmer

Reason

Potential risk of electrical fire when fluid leaks into the power receptacle on the back of the warmer. This risk exists any time the warmer is plugged in.

Affected products

NxStage System One Express Fluid Warmer

Lot or serial number

All lots.

Model or catalog number

FW-300-A

Companies
Manufacturer
NxStage Medical, Inc.
350 Merrimack Street
Lawrence
01843
Massachusetts
UNITED STATES