This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

NUCLISENS LYSIS BUFFER (2017-07-13)

Starting date:
July 13, 2017
Posting date:
October 30, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Hospitals, Healthcare Professionals
Identification number:
RA-64990

Affected products

NUCLISENS LYSIS BUFFER

Reason

Following a customer complaint related to coloured eluates for whole blood extractions when using NucliSENS Lysis Buffer ref 200292 - lot 17022802, an R&D investigation showed that the eluate becomes coloured because of the presence of the group heme.

Affected products

NUCLISENS LYSIS BUFFER

Lot or serial number

17022802

Model or catalog number

200292

Companies
Manufacturer

Biomerieux Sa

376 Chemin De L'orme

Marcy-l'etoile

69280

FRANCE