This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

NUCLISENS EASYMAG - MAGNETIC SILICA (2016-07-08)

Starting date:
July 8, 2016
Posting date:
July 19, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type I
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-59440

Affected products

A. NUCLISENS EASYMAG - MAGNETIC SILICA

Reason

A decrease of extraction performance was observed for some specific downstream applications due to the magnetic silica (Magsil) (ref:280133). The decrease of downstream application performances can lead to a risk of false negative, invalid or under-quantification results.

Affected products

A. NUCLISENS EASYMAG - MAGNETIC SILICA

Lot or serial number

Z017CD1MS

Z017CG1MS

Z017CH1MS

Model or catalog number

280133

Companies
Manufacturer

Biomerieux Sa

Chemin De L'orme

Marcy-l'etoile

69280

FRANCE