Nucleus 7 Processing Unit
Brand(s)
Summary
Acknowledge receipt of notice, contact Cochlear rep to manage recalled units and arrange replacement.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Nucleus 7 Processing Unit |
More than 10 numbers, contact manufacturer. |
Z482008 |
Issue
This failure mode has potential to manifest for all devices manufactured with apple MFi keys in the range 6100 to 6103 when attempted to pair with iOS devices. Complaints to date have come from CAM, APAC and EMEA regions. This issue is isolated to cp1000 and cp1001 devices manufactured between 1st April to 19th May 2022 when used with any device using Apple iOS.
Recall start date: June 3, 2022
Additional information
Details
Cochlear Limited
1 University Avenue, Macquarie Park, New South Wales, Australia, 2109
Get notified