This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Novopen 5 and Novopen Echo (2017-07-05)

Starting date:
July 5, 2017
Posting date:
July 18, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-64024

Affected Products

A. Novopen 5

B. Novopen Echo

Reason

Novo Nordisk A/S has detected that the insulin cartridge holder used in a select number of NovoPen Echo and NovoPen 5 lots may crack or break if exposed to certain chemicals, for example certain cleaning agents. NovoPen Echo and NovoPen 5 are used for insulin treatment by people living with diabetes. NovoPen 4 and FlexTouch are not impacted.

Affected products

A. Novopen 5

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

0 59014 00520 4

0 59014 00522 8

Companies
Manufacturer

Novo Nordisk A/S Novo Nordisk (China) Pharmaceuticals Co. Ltd.

No 99 Nanhai Road

Tianjin

300457

CHINA



B. Novopen Echo

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

0 59014 00441 2

0 59014 00442 9

Companies
Manufacturer

Novo Nordisk A/S Novo Nordisk (China) Pharmaceuticals Co. Ltd.

No 99 Nanhai Road

Tianjin

300457

CHINA



Important safety information for people with diabetes using NovoPen Echo or NovoPen 5

2017-07-25 | Health products

Advisory