This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
Novocastra Lyophilized Mouse Monoclonal Antibody CD10 (2015-09-03)
- Starting date:
- September 3, 2015
- Posting date:
- October 5, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-55204
Affected Products
Novocastra Lyophilized Mouse Monoclonal Antibody CD10
Reason
The manufacturer informed Leica Microsystems that testing of retained units, as part of an ongoing stability program, has shown that lot numbers 6033086, 6033622 and 6032742 are not stable up to the expiry date on the product labelling. Only lot number 6032742 has been distributed in Canada
Affected products
Novocastra Lyophilized Mouse Monoclonal Antibody CD10
Lot or serial number
6032742
Model or catalog number
NCL-CD10-270
Companies
- Manufacturer
-
Leica Biosystems Newcastle Ltd.
Balliol Business Park West
Benton Lane
UNITED KINGDOM