Novastar Full Face Mask
Brand(s)
Last updated
Summary
Product
Novastar Full Face Mask
Issue
Medical devices - Performance issue
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Full Face Mask Novastar Ts, Niv, With Anti-Asphyxia Valve |
All lots. |
MP01576-XX |
Full Face Mask Novastar Ts, Niv, With Standard Elbow |
All lots. |
MP01580-XX |
Issue
In case the magnets of the Novastar TS and Novastar Plus are positioned near medical devices (implanted as well as not-implanted) and metallic implants the magnetic field might impact the implants or devices and cause serious injury.
Recall start date: January 13, 2023
Additional information
Details
Original published date: 2023-01-24
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies
Draegerwerk Ag & Co. Kgaa
Moislinger Allee 53-55, Lubeck, Germany, 23542
Published by
Health Canada
Audience
Healthcare
Recall class
Type I
Identification number
RA-72611