NovaGuide™ Trditional and Intelligent Ultrasound Systems
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
NovaGuide™ 2 Intelligent Ultrasound System |
Serial number 23023421 |
NSC-NVGSYS2-CA |
NovaGuide™ 2 Traditional Ultrasound System |
Serial number 22074378 |
NSC-TCDNG2 |
Issue
A potential for an unintentional disconnection of the internal chasis ground cable was identified. This was initially addressed on 02/16/24, with a final design on 04/29/24. The cable is not accessible during normal operation and the inadvertant disconnection is a remote possibility.
Recall start date: May 6, 2024
Additional information
Details
Neuralsignal Inc Dba. Novasignal Corp.
1109 Westwood Blvd., Los Angeles, California, United States, 90024
Get notified
Receive notifications for new and updated recalls and alerts by category.