Non Absorbable Surgical Suture Stainless Steel (2014-03-24)

Starting date:
March 24, 2014
Posting date:
April 15, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-39081

Recalled products

  1. Non Absorbable Surgical Suture Stainless Steel

Reason

Certain lots of Teleflex sutures did not meet minimum diameter requirements. If affected product is used and failure occurs, delay in procedure, wound dehiscence, bleeding and/or infection may occur requiring medical and/or surgical intervention.

Affected products

A. Non Absorbable Surgical Suture Stainless Steel

Lot or serial number
  • 02D1100604
  • 02K1100784
  • 02M1101954
Model or catalog number
  • 29 727
  • 29-7171M4
  • 29-7270M4
Companies
Manufacturer

Teleflex Medical

2917 Wreck Drive,

Research Triangle Park

27709

North Carolina