Noblus Diagnostic Ultrasound Scanner - Main Unit
Brand(s)
Summary
Contact the importer CHRISTIEINNOMED INC at QUALITE.QUALITY@CHRISTIEINNOMED.COM.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Noblus Diagnostic Ultrasound Scanner - Main Unit |
K1716 0000 |
NOBLUS SYSTEM |
Issue
Fujifilm has become aware that some of the AC adapters of Noblus heat up and cause smoke, in the worst case, when excessive mechanical stress is applied to the socket for power cable (frequent insertion/removal of the power cable). This mechanical stress would cause failure of the solder joint inside of the AC adapter and generates excessive heat.
Recall Start Date: May 9, 2024
Additional information
Details
Fujifilm Healthcare Corporation
2-1 Shintoyofuta, Kashiwa-Shi, Chiba-Ken, Japan, 277-0804