Health product recall

NobelProcera Full Contour Zirconia (FCZ) Implant Crown (2021-04-09)

Starting date:
April 9, 2021
Posting date:
July 7, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-75977

Last updated: 2021-07-07

Affected Products 

NobelProcera Full Contour Zirconia (FCZ) Implant Crown

Reason

Devices were not produced according to specifications. The sintering furnace did not execute the correct sintering program, and the affected devices did therefore not complete the entire sintering procedure.

Affected products

NobelProcera Full Contour Zirconia (FCZ) Implant Crown

Lot or serial number

More than 10 numbers, contact manufacturer. 

Model or catalog number
  • 77352
  • 77363
Companies
Manufacturer
Nobel Biocare AB
Vastra Hamngatan 1
GOTEBORG
40226
SWEDEN