Nitric Oxide Therapy Adapters And Connectors
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Nitric Oxide Therapy Adapters And Connectors |
74M2001125 |
50412 |
Issue
A single lot (74M2001125) of the one-way valve (part #50412 (pack of 5)) utilized with the Inomax dsir plus is being recalled. The one-way valve is used to prevent reverse flow into the Inomax dsir injector module from high frequency ventilators and spontaneous mask setups. The recall is being initiated due to the possibility of this lot of one-way valves sticking, by preventing or reducing the breathing circuit flow. We have no reports of serious injuries due to this issue.
Recall Start Date: August 8, 2023
Additional information
Details
Mallinckrodt Manufacturing Llc
6603 Femrite Drive, Madison, Wisconsin, United States, 53718
Get notified