Health product recall

Nitric Oxide Therapy Adapters And Connectors

Last updated

Summary

Product
Nitric Oxide Therapy Adapters And Connectors
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Affected products

Affected products

Lot or serial number

Model or catalog number

Nitric Oxide Therapy Adapters And Connectors

74M2001125

50412

Issue

A single lot (74M2001125) of the one-way valve (part #50412 (pack of 5)) utilized with the Inomax dsir plus is being recalled. The one-way valve is used to prevent reverse flow into the Inomax dsir injector module from high frequency ventilators and spontaneous mask setups. The recall is being initiated due to the possibility of this lot of one-way valves sticking, by preventing or reducing the breathing circuit flow. We have no reports of serious injuries due to this issue.

Recall Start Date: August 8, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies

Mallinckrodt Manufacturing Llc

6603 Femrite Drive, Madison, Wisconsin, United States, 53718

Published by
Health Canada
Audience
General public
Healthcare
Industry
Recall class
Type II
Identification number
RA-74141

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe