Health product recall

NextSeq™ 550Dx and MiSeq™Dx Instruments

Brand(s)
Last updated

Summary

Product
NextSeq™ 550Dx and MiSeq™Dx Instruments
Issue
Medical devices - Performance issue
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

NextSeq™ 550Dx Instrument

NDX551020
NDX551031

20005715

MiSeq™Dx Instrument

M70349
M70553
M70552
M70444
M70440
M70546
M70335

DX-410-1001

Issue

A concern with the software network binding of the Universal Copy  Service (UCS), an ILMN proprietary software component present on ILMN instruments. The concern expressed was that the software process running as a tcp service was found to bind to all interfaces ("*") in UCS V.2.X*.

Recall start date: April 5, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Health products - Medical devices - Gastroenterology & Urology
Companies

Illumina, Inc.

5200 Illumina Way, San Diego, California, United States, 92122

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-73736

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe