NextSeq™ 550Dx and MiSeq™Dx Instruments
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
NextSeq™ 550Dx Instrument |
NDX551020 |
20005715 |
MiSeq™Dx Instrument |
M70349 |
DX-410-1001 |
Issue
A concern with the software network binding of the Universal Copy Service (UCS), an ILMN proprietary software component present on ILMN instruments. The concern expressed was that the software process running as a tcp service was found to bind to all interfaces ("*") in UCS V.2.X*.
Recall start date: April 5, 2023
Additional information
Details
Illumina, Inc.
5200 Illumina Way, San Diego, California, United States, 92122