Neuromate Stereotactic System
Brand(s)
Last updated
Summary
Product
Neuromate Stereotactic System
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Neuromate Stereotactic System |
All lots. |
001.0050 |
Issue
A bug has been detected in the planning software. When loading images, there may be a left/right inversion. Display of image series is flipped after importing a sagittal oriented image series stored using the enhanced Dicom format.
Recall start date: July 8, 2022
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Neurology
Companies
Renishaw Mayfield Sarl
31 Rue Ampere, Chassieu, Rhone, France, 69680
Published by
Health Canada
Audience
Healthcare
Recall class
Type I
Identification number
RA-64374
Get notified
Receive emails about new and updated recall and safety alerts.