NEPTUNE 2 ROVER - DIVERTER VALVE ASSEMBLY Component (2015-06-09)

Starting date:
June 9, 2015
Posting date:
June 26, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Identification number:
RA-53954

Recalled Product

A.NEPTUNE 2 ROVER - DIVERTER VALVE ASSEMBLY Component

Reason

It has been identified during the installation of rovers, there is a potential for the Neptune 2 Rover's diverter valve assembly to not actuate during docking, preventing the cleaning cycle from occurring and inhibiting prefill water from being supplied to the Neptune fresh water system. The diverter valve's material combined with the heat-treating process has the potential to cause diverter valve corrosion when the rover is allowed to sit for a period of greater than 40 days.

Affected products

A.NEPTUNE 2 ROVER - DIVERTER VALVE ASSEMBLY Component

Lot or serial number

0803118093 to 1505400223

Model or catalog number

0702-001-000

Companies
Manufacturer

Stryker Instruments

4100 East Milham Ave

Kalamazoo

Michigan

UNITED STATES