NEOFEED ENTERAL SYRINGE PUMP (2021-08-17)
- Starting date:
- August 17, 2021
- Posting date:
- September 24, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-76495
Last updated:
Affected Products
NEOFEED ENTERAL SYRINGE PUMP
Reason
CME has identified that the syringe pump motor block mechanism may be affected overtime by "wear / tear". This scenario causes under infusion situations without the pump alarm being generated. The issue can be observed when the actuator does not move when the lead screw is rotating, resulting in potentially no pressure on syringe plunger to deliver the syringe contents.
Affected products
NEOFEED ENTERAL SYRINGE PUMP
Lot or serial number
All lots
Model or catalog number
NEOFEED
Companies
- Manufacturer
-
Caesarea Medical Electronics Ltd.
16 Shacham Street, Industrial Park P. O. Box 4294
Caesarea
38900
ISRAEL