neoBLUE blanket LED Phototherapy System (2019-01-08)
- Starting date:
- January 8, 2019
- Posting date:
- February 8, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-68986
Affected Products
neoBLUE blanket LED Phototherapy System
Reason
This notification is related to a prior notification, October 2016, regarding the neoBLUE blanket LED Phototherapy System. At that time, Natus Medical informed you of a potential issue of discoloration/degradation/melting of the fibre optic bundle at the pad connector which is inserted in the neoBLUE blanket light box. Natus Medical completed a redesign of the device and is initiating this field action to replace those systems shipped from 2011 through march 21, 2016. Newer systems shipped after March 21, 2016 do not need to be replaced. Natus will provide an updated literature kit for each of the newer devices.
Affected products
neoBLUE blanket LED Phototherapy System
Lot or serial number
- More than 10 numbers, contact manufacturer.
Model or catalog number
- 006244
- 006895
Companies
- Manufacturer
-
Natus Medical Incorporated
5900 First Avenue South
Seattle, Washington
98108
UNITED STATES