This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

NEOBASE SUCCINYLACETONE ASSAY SOLUTION (2017-11-02)

Starting date:
November 2, 2017
Posting date:
November 22, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Healthcare Professionals, Hospitals, General Public
Identification number:
RA-65206

Affected products

NEOBASE SUCCINYLACETONE ASSAY SOLUTION

Reason

The vial and stopper used in this assay solution is provided by an external supplier. It was observed, during manufacturing of a lot, that some of the stoppers within the reagent vials were assembled incorrectly by the supplier. The seal was deformed or assembled upside down. Two of the affected vials were leaking.

Affected products

NEOBASE SUCCINYLACETONE ASSAY SOLUTION

Lot or serial number

654184
655300

Model or catalog number

3042-0020

Companies
Manufacturer
Wallac Oy
Mustionkatu 6, P.O. Box 10
Turku
20750
FINLAND