Health product recall

Neo-Tee T-Piece Resuscitator

Brand(s)
Last updated

Summary

Product
Neo-Tee T-Piece Resuscitator
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Industry

Affected products

Affected products Lot or serial number Model or catalogue number
Neo-Tee T-Piece Resuscitator 2426450809 10-50809

Issue

To address the potential for performance specifications not being met. The device may not reach the high and the low ends of the peak inspiration pressure (PIP) and positive end expiratory pressure (PEEP) ranges as a result of an undersized spring in the controller.

Recall start date: December 5, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies
Mercury Medical
11300 49th St. N., Clearwater, Florida, United States, 33762
Published by
Health Canada
Audience
Industry
Recall class
Type I
Identification number
RA-76755

Get notified

Receive notifications for new and updated recalls and alerts by category.