This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

NEO-TEE (T-PIECE RESUSCITATOR) (2015-11-17)

Starting date:
November 17, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-56192

Affected Products

NEO-TEE (T-PIECE RESUSCITATOR)

Reason

Potential of a silicone mask with a 15 mm male port being able to be forcibly inserted into the protective cap covering the opening of the T-piece Circuit. Incorrect use of the product in this way will result in complete occlusion of the patient circuit and consequent failure to ventilate the patient resulting in hypoxia or delay in patient care.

Affected products

NEO-TEE (T-PIECE RESUSCITATOR)

Lot or serial number

All lots.

Model or catalog number
  • M1091316VS
  • M1091316VS
  • M1091335VS
  • M1091335VS
  • M1091365VS
  • M1091365VS
Companies
Manufacturer

Mercury Medical

11300 49TH ST. N.

Clearwater

33762

Florida

UNITED STATES