Neo-Tee T-Piece Resuscitator
Brand(s)
Last updated
Summary
Product
Neo-Tee T-Piece Resuscitator
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Industry
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Neo-Tee T-Piece Resuscitator |
2426450809 |
10-50809 |
Issue
To address the potential for performance specifications not being met. The device may not reach the high and the low ends of the peak inspiration pressure (PIP) and positive end expiratory pressure (PEEP) ranges as a result of an undersized spring in the controller.
Recall start date: December 5, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies
| Mercury Medical |
| 11300 49th St. N., Clearwater, Florida, United States, 33762 |
Published by
Health Canada
Audience
Industry
Recall class
Type I
Identification number
RA-76755
Get notified
Receive emails about new and updated recall and safety alerts.