Health product recall

Neo-Fit Neonatal Endotracheal Tube Grip

Last updated

Summary

Product
Neo-Fit Neonatal Endotracheal Tube Grip
Issue
Medical devices - Performance issue
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
Neo-Fit Neonatal Endotracheal Tube Grip All lots. 42-2540

Issue

Coopersurgical has become aware of a potential health risk in which metal clips used to grip the neonatal endotracheal tube may fracture and detach from the strap and then be ingested by the patient. If detached metal clips are ingested, there is a potential for laceration to the gastrointestinal tract as the clip passes through the patient's body.

Recall start date: October 18, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Ear, nose and throat
Companies

Coopersurgical Inc.

95 Corporate Drive, Trumbull, Connecticut, United States, 06611

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-72176

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe