Health product recall

Negative Pressure Wound Therapy Sved Device

Brand(s)
Last updated

Summary

Product
Negative Pressure Wound Therapy Sved Device
Issue
Medical devices - Revised instructions for use
What to do

Contact the manufacturer if you require additional information.

Affected products

Affected Products

Lot or Serial Number

Model or catalog number

Negative Pressure Wound Therapy Sved Device

All lots.

6701132BL, 6701132BL-DE, 6701132, 6701132BL-A

Issue

Cardinal Health is initiating an urgent medical device correction to update the clinician and patient user manuals for the Negative Pressure Wound Therapy Sved Device to add pertinent information related to periodic preventative maintenance for this device. The current three year service life stated in the clinician and patient user manuals has been augmented to include a requirement to service the units every 3 years or once the unit has reached 75 days of runtime, whichever comes first.

Recall start date: Apr 19, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Cardinal Health

3651 Birchwood Drive, Waukegan, Illinois

United States, 60085

Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type II
Identification number
RA-64113

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe