Negative Pressure Wound Therapy Sved Device
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected Products |
Lot or Serial Number |
Model or catalog number |
---|---|---|
Negative Pressure Wound Therapy Sved Device |
All lots. |
6701132BL, 6701132BL-DE, 6701132, 6701132BL-A |
Issue
Cardinal Health is initiating an urgent medical device correction to update the clinician and patient user manuals for the Negative Pressure Wound Therapy Sved Device to add pertinent information related to periodic preventative maintenance for this device. The current three year service life stated in the clinician and patient user manuals has been augmented to include a requirement to service the units every 3 years or once the unit has reached 75 days of runtime, whichever comes first.
Recall start date: Apr 19, 2022
Additional information
Details
Cardinal Health
3651 Birchwood Drive, Waukegan, Illinois
United States, 60085