Health product recall

NAVIGUS STIMLOC (2021-06-04)



Starting date:
June 4, 2021
Posting date:
June 30, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-75925


Last updated:

Affected Products 

NAVIGUS STIMLOC

Reason

Medtronic has identified that specific lots of the Stimloc Clips may have excess nylon molding flash. This issue can prevent the clip from snapping into the StimLoc Base, which could allow for lead movement or lead damage. This could potentially result in delivery of therapy to the wrong location, inadequate delivery of therapy, or intracranial hemorrhage.

Affected products

NAVIGUS STIMLOC

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

924256

Companies
Manufacturer

Medtronic Inc.