Health product recall

NATRELLE INSPIRA TRL, TRLP, TRM, TRF, TRX TruForm 1 (responsive), NATRELLE INSPIRA TSL,TSLP, TSM, TSF, TSX TruForm 2 (soft touch), NATRELLE 410FL, 410ML, 410LL, 410FM, 410MM, 410LM, 410FF, 410MF, 410LF, 410FX, 410MX, 410LX TruForm Silicone -Filled Br...

Starting date:
May 30, 2019
Posting date:
June 7, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-70139

Last updated: 2019-06-07

Affected Products

A. NATRELLE INSPIRA TRL TruForm 1 (responsive)
B. NATRELLE INSPIRA TRLP TruForm 1 (responsive)
C. NATRELLE INSPIRA TRM TruForm 1 (responsive)
D. NATRELLE INSPIRA TRF TruForm 1 (responsive)
E. NATRELLE INSPIRA TRX TruForm 1 (responsive)
F. NATRELLE INSPIRA TSL TruForm 2 (soft touch)
G. NATRELLE INSPIRA TSLP TruForm 2 (soft touch)
H. NATRELLE INSPIRA TSM TruForm 2 (soft touch)
I. NATRELLE INSPIRA TSF TruForm 2 (soft touch)
J. NATRELLE INSPIRA TSX TruForm 2 (soft touch)
K. NATRELLE 410FL TruForm Silicone - Filled Breast Implant
L. NATRELLE 410ML TruForm Silicone - Filled Breast Implant
M. NATRELLE 410LL TruForm Silicone - Filled Breast Implant
N. NATRELLE 410FM TruForm Silicone - Filled Breast Implant
O. NATRELLE 410MM TruForm Silicone - Filled Breast Implant
P. NATRELLE 410LM TruForm Silicone - Filled Breast Implant
Q. NATRELLE 410FF TruForm Silicone - Filled Breast Implant
R. NATRELLE 410MF TruForm Silicone - Filled Breast Implant
S. NATRELLE 410LF TruForm Silicone - Filled Breast Implant
T. NATRELLE 410FX TruForm Silicone - Filled Breast Implant
U. NATRELLE 410MX TruForm Silicone - Filled Breast Implant
V. NATRELLE 410LX TruForm Silicone - Filled Breast Implant
W. NATRELLE 410FF TruForm Soft Touch Silicone - Filled Breast Implant
X. NATRELLE 410FL TruForm Soft Touch Silicone - Filled Breast Implant
Y. NATRELLE 410FM TruForm Soft Touch Silicone - Filled Breast Implant
Z. NATRELLE 410LF TruForm Soft Touch Silicone - Filled Breast Implant
AA. NATRELLE 410LL TruForm Soft Touch Silicone - Filled Breast Implant
AB. NATRELLE 410LM TruForm Soft Touch Silicone - Filled Breast Implant
AC. NATRELLE 410MF TruForm Soft Touch Silicone - Filled Breast Implant
AD. NATRELLE 410ML TruForm Soft Touch Silicone - Filled Breast Implant
AE. NATRELLE 410MM TruForm Soft Touch Silicone - Filled Breast Implant

Reason

Health Canada has suspended the licenses for Allergan's Biocell textured breast implants (the only macro-textured implants available in Canada), based upon the rare but serious risk of breast implant-associated anaplastic large cell lymphoma (BIA-ALCL), Allergan is voluntarily recalling BioCell breast implants as a precautionary measure.

Affected products

A. NATRELLE INSPIRA TRL TruForm 1 (responsive)

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

N-TRL110
N-TRL125
N-TRL140
N-TRL170
N-TRL200
N-TRL230
N-TRL260
N-TRL290
N-TRL320
N-TRL350
N-TRL380
N-TRL410
N-TRL460
N-TRL510
N-TRL560
N-TRL610

Companies
Manufacturer
ALLERGAN
MARLOW INTERNATIONAL, PARKWAY
MARLOW, BUCKINGHAMSHIRE,
SL7 1YL
UNITED KINGDOM

B. NATRELLE INSPIRA TRLP TruForm 1 (responsive)

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

N-TRLP125
N-TRLP145
N-TRLP165
N-TRLP190
N-TRLP205
N-TRLP220
N-TRLP235
N-TRLP250
N-TRLP265
N-TRLP280
N-TRLP300
N-TRLP320
N-TRLP340
N-TRLP360
N-TRLP400
N-TRLP440
N-TRLP490
N-TRLP540
N-TRLP590
N-TRLP640

Companies
Manufacturer
ALLERGAN
MARLOW INTERNATIONAL, PARKWAY
MARLOW, BUCKINGHAMSHIRE,
SL7 1YL
UNITED KINGDOM

C. NATRELLE INSPIRA TRM TruForm 1 (responsive)

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

N-TRM140
N-TRM175
N-TRM195
N-TRM210
N-TRM240
N-TRM255
N-TRM275
N-TRM295
N-TRM310
N-TRM330
N-TRM345
N-TRM360
N-TRM375
N-TRM405
N-TRM445
N-TRM485
N-TRM520
N-TRM560
N-TRM600
N-TRM640
N-TRM685
N-TRM755

Companies
Manufacturer
ALLERGAN
MARLOW INTERNATIONAL, PARKWAY
MARLOW, BUCKINGHAMSHIRE,
SL7 1YL
UNITED KINGDOM

D. NATRELLE INSPIRA TRF TruForm 1 (responsive)

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

N-TRF180
N-TRF200
N-TRF220
N-TRF240
N-TRF265
N-TRF295
N-TRF325
N-TRF335
N-TRF345
N-TRF365
N-TRF385
N-TRF415
N-TRF450
N-TRF485
N-TRF520
N-TRF560
N-TRF605
N-TRF650
N-TRF695
N-TRF745
N-TRF770

Companies
Manufacturer
ALLERGAN
MARLOW INTERNATIONAL, PARKWAY
MARLOW, BUCKINGHAMSHIRE,
SL7 1YL
UNITED KINGDOM

E. NATRELLE INSPIRA TRX TruForm 1 (responsive)

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

N-TRX205
N-TRX225
N-TRX255
N-TRX285
N-TRX310
N-TRX340
N-TRX375
N-TRX400
N-TRX420
N-TRX445
N-TRX470
N-TRX495
N-TRX525
N-TRX545
N-TRX560
N-TRX580
N-TRX615
N-TRX650
N-TRX700
N-TRX750
N-TRX800

Companies
Manufacturer
ALLERGAN
MARLOW INTERNATIONAL, PARKWAY
MARLOW, BUCKINGHAMSHIRE,
SL7 1YL
UNITED KINGDOM

F. NATRELLE INSPIRA TSL TruForm 2 (soft touch)

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

N-TSL110
N-TSL125
N-TSL140
N-TSL170
N-TSL200
N-TSL230
N-TSL260
N-TSL290
N-TSL320
N-TSL350
N-TSL380
N-TSL410
N-TSL460
N-TSL510
N-TSL560
N-TSL610

Companies
Manufacturer
ALLERGAN
MARLOW INTERNATIONAL, PARKWAY
MARLOW, BUCKINGHAMSHIRE,
SL7 1YL
UNITED KINGDOM

G. NATRELLE INSPIRA TSLP TruForm 2 (soft touch)

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

N-TSLP125
N-TSLP145
N-TSLP165
N-TSLP190
N-TSLP205
N-TSLP220
N-TSLP235
N-TSLP250
N-TSLP265
N-TSLP280
N-TSLP300
N-TSLP320
N-TSLP340
N-TSLP360
N-TSLP400
N-TSLP440
N-TSLP490
N-TSLP540
N-TSLP590
N-TSLP640

Companies
Manufacturer
ALLERGAN
MARLOW INTERNATIONAL, PARKWAY
MARLOW, BUCKINGHAMSHIRE,
SL7 1YL
UNITED KINGDOM

H. NATRELLE INSPIRA TSM TruForm 2 (soft touch)

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

N-TSM140
N-TSM175
N-TSM195
N-TSM210
N-TSM240
N-TSM255
N-TSM275
N-TSM295
N-TSM310
N-TSM330
N-TSM345
N-TSM360
N-TSM375
N-TSM405
N-TSM445
N-TSM485
N-TSM520
N-TSM560
N-TSM600
N-TSM640
N-TSM685
S-TSM755

Companies
Manufacturer
ALLERGAN
MARLOW INTERNATIONAL, PARKWAY
MARLOW, BUCKINGHAMSHIRE,
SL7 1YL
UNITED KINGDOM

I. NATRELLE INSPIRA TSF TruForm 2 (soft touch)

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

N-TSF180
N-TSF200
N-TSF220
N-TSF240
N-TSF265
N-TSF295
N-TSF325
N-TSF335
N-TSF345
N-TSF365
N-TSF385
N-TSF415
N-TSF450
N-TSF485
N-TSF520
N-TSF560
N-TSF605
N-TSF650
N-TSF695
N-TSF745
N-TSF770

Companies
Manufacturer
ALLERGAN
MARLOW INTERNATIONAL, PARKWAY
MARLOW, BUCKINGHAMSHIRE,
SL7 1YL
UNITED KINGDOM

J. NATRELLE INSPIRA TSX TruForm 2 (soft touch)

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

N-TSX205
N-TSX225
N-TSX255
N-TSX285
N-TSX310
N-TSX340
N-TSX375
N-TSX400
N-TSX420
N-TSX445
N-TSX470
N-TSX495
N-TSX525
N-TSX545
N-TSX560
N-TSX580
N-TSX615
N-TSX650
N-TSX700
N-TSX750
N-TSX800

Companies
Manufacturer
ALLERGAN
MARLOW INTERNATIONAL, PARKWAY
MARLOW, BUCKINGHAMSHIRE,
SL7 1YL
UNITED KINGDOM

K. NATRELLE 410FL TruForm Silicone - Filled Breast Implant

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

N-27-FL095-120
N-27-FL100-140
N-27-FL105-170
N-27-FL110-190
N-27-FL115-220
N-27-FL120-250
N-27-FL125-285
N-27-FL130-320
N-27-FL135-360
N-27-FL140-400
N-27-FL145-440
N-27-FL150-495

Companies
Manufacturer
ALLERGAN
MARLOW INTERNATIONAL, PARKWAY
MARLOW, BUCKINGHAMSHIRE,
SL7 1YL
UNITED KINGDOM

L. NATRELLE 410ML TruForm Silicone - Filled Breast Implant

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

N-27-ML095-110
N-27-ML100-125
N-27-ML105-150
N-27-ML110-170
N-27-ML115-195
N-27-ML120-220
N-27-ML125-250
N-27-ML130-285
N-27-ML135-320
N-27-ML140-360
N-27-ML145-390
N-27-ML150-440

Companies
Manufacturer
ALLERGAN
MARLOW INTERNATIONAL, PARKWAY
MARLOW, BUCKINGHAMSHIRE,
SL7 1YL
UNITED KINGDOM

M. NATRELLE 410LL TruForm Silicone - Filled Breast Implant

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

N-27-LL095-100
N-27-LL100-115
N-27-LL105-135
N-27-LL110-160
N-27-LL115-180
N-27-LL120-210
N-27-LL125-240
N-27-LL130-270
N-27-LL135-300
N-27-LL140-340
N-27-LL145-370
N-27-LL150-415
N-27-LL155-455

Companies
Manufacturer
ALLERGAN
MARLOW INTERNATIONAL, PARKWAY
MARLOW, BUCKINGHAMSHIRE,
SL7 1YL
UNITED KINGDOM

N. NATRELLE 410FM TruForm Silicone - Filled Breast Implant

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

N-27-FM095-155
N-27-FM100-180
N-27-FM105-205
N-27-FM110-235
N-27-FM115-270
N-27-FM120-310
N-27-FM125-350
N-27-FM130-395
N-27-FM135-440
N-27-FM140-500
N-27-FM145-550
N-27-FM150-605
N-27-FM155-670

Companies
Manufacturer
ALLERGAN
MARLOW INTERNATIONAL, PARKWAY
MARLOW, BUCKINGHAMSHIRE,
SL7 1YL
UNITED KINGDOM

O. NATRELLE 410MM TruForm Silicone - Filled Breast Implant

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

N-27-MM095-135
N-27-MM100-160
N-27-MM105-185
N-27-MM110-215
N-27-MM115-245
N-27-MM120-280
N-27-MM125-320
N-27-MM130-360
N-27-MM135-400
N-27-MM140-450
N-27-MM145-500
N-27-MM150-555

Companies
Manufacturer
ALLERGAN
MARLOW INTERNATIONAL, PARKWAY
MARLOW, BUCKINGHAMSHIRE,
SL7 1YL
UNITED KINGDOM

P. NATRELLE 410LM TruForm Silicone - Filled Breast Implant

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

N-27-LM095-115
N-27-LM100-140
N-27-LM105-160
N-27-LM110-190
N-27-LM115-220
N-27-LM120-250
N-27-LM125-285
N-27-LM130-320
N-27-LM135-360
N-27-LM140-400
N-27-LM145-450
N-27-LM150-495
N-27-LM155-545

Companies
Manufacturer
ALLERGAN
MARLOW INTERNATIONAL, PARKWAY
MARLOW, BUCKINGHAMSHIRE,
SL7 1YL
UNITED KINGDOM

Q. NATRELLE 410FF TruForm Silicone - Filled Breast Implant

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

N-27-FF095-160
N-27-FF100-185
N-27-FF105-220
N-27-FF110-255
N-27-FF115-290
N-27-FF120-335
N-27-FF125-375
N-27-FF130-425
N-27-FF135-475
N-27-FF140-535
N-27-FF145-595
N-27-FF150-655
N-27-FF155-740

Companies
Manufacturer
ALLERGAN
MARLOW INTERNATIONAL, PARKWAY
MARLOW, BUCKINGHAMSHIRE,
SL7 1YL
UNITED KINGDOM

R. NATRELLE 410MF TruForm Silicone - Filled Breast Implant

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

N-27-MF095-140
N-27-MF100-165
N-27-MF105-195
N-27-MF110-225
N-27-MF115-255
N-27-MF120-295
N-27-MF125-335
N-27-MF130-375
N-27-MF135-420
N-27-MF140-470
N-27-MF145-525
N-27-MF150-580
N-27-MF155-640

Companies
Manufacturer
ALLERGAN
MARLOW INTERNATIONAL, PARKWAY
MARLOW, BUCKINGHAMSHIRE,
SL7 1YL
UNITED KINGDOM

S. NATRELLE 410LF TruForm Silicone - Filled Breast Implant

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

N-27-LF095-125
N-27-LF100-150
N-27-LF105-175
N-27-LF110-205
N-27-LF115-240
N-27-LF120-270
N-27-LF125-310
N-27-LF130-350
N-27-LF135-390
N-27-LF140-440
N-27-LF145-490
N-27-LF150-540
N-27-LF155-595

Companies
Manufacturer
ALLERGAN
MARLOW INTERNATIONAL, PARKWAY
MARLOW, BUCKINGHAMSHIRE,
SL7 1YL
UNITED KINGDOM

T. NATRELLE 410FX TruForm Silicone - Filled Breast Implant

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

N-27-FX095-185
N-27-FX100-215
N-27-FX105-245
N-27-FX110-280
N-27-FX115-315
N-27-FX120-360
N-27-FX125-410
N-27-FX130-450
N-27-FX135-495
N-27-FX140-560
N-27-FX145-615
N-27-FX150-690
N-27-FX155-775

Companies
Manufacturer
ALLERGAN
MARLOW INTERNATIONAL, PARKWAY
MARLOW, BUCKINGHAMSHIRE,
SL7 1YL
UNITED KINGDOM

U. NATRELLE 410MX TruForm Silicone - Filled Breast Implant

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

N-27-MX095-165
N-27-MX100-195
N-27-MX105-225
N-27-MX110-255
N-27-MX115-290
N-27-MX120-325
N-27-MX125-370
N-27-MX130-410
N-27-MX135-445
N-27-MX140-520
N-27-MX145-550
N-27-MX150-620
N-27-MX155-685
N-27-MX160-765

Companies
Manufacturer
ALLERGAN
MARLOW INTERNATIONAL, PARKWAY
MARLOW, BUCKINGHAMSHIRE,
SL7 1YL
UNITED KINGDOM

V. NATRELLE 410LX TruForm Silicone - Filled Breast Implant

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

N-27-LX095-145
N-27-LX100-175
N-27-LX105-195
N-27-LX110-225
N-27-LX115-255
N-27-LX120-290
N-27-LX125-330
N-27-LX130-365
N-27-LX135-405
N-27-LX140-455
N-27-LX145-515
N-27-LX150-570
N-27-LX155-625
N-27-LX160-685

Companies
Manufacturer
ALLERGAN
MARLOW INTERNATIONAL, PARKWAY
MARLOW, BUCKINGHAMSHIRE,
SL7 1YL
UNITED KINGDOM

W. NATRELLE 410FF TruForm Soft Touch Silicone - Filled Breast Implant

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

N-ST-FF095-160
N-ST-FF100-185
N-ST-FF105-220
N-ST-FF110-255
N-ST-FF115-290
N-ST-FF120-335
N-ST-FF125-375
N-ST-FF130-425
N-ST-FF135-475
N-ST-FF140-535
N-ST-FF145-595
N-ST-FF150-655
N-ST-FF155-740

Companies
Manufacturer
ALLERGAN
MARLOW INTERNATIONAL, PARKWAY
MARLOW, BUCKINGHAMSHIRE,
SL7 1YL
UNITED KINGDOM

X. NATRELLE 410FL TruForm Soft Touch Silicone - Filled Breast Implant

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

N-ST-FL100-140
N-ST-FL110-190
N-ST-FL115-220
N-ST-FL120-250
N-ST-FL130-320

Companies
Manufacturer
ALLERGAN
MARLOW INTERNATIONAL, PARKWAY
MARLOW, BUCKINGHAMSHIRE,
SL7 1YL
UNITED KINGDOM

Y. NATRELLE 410FM TruForm Soft Touch Silicone - Filled Breast Implant

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

N-ST-FM095-155
N-ST-FM100-180
N-ST-FM105-205
N-ST-FM110-235
N-ST-FM115-270
N-ST-FM120-310
N-ST-FM125-350
N-ST-FM130-395
N-ST-FM135-440
N-ST-FM140-500
N-ST-FM145-550
N-ST-FM150-605
N-ST-FM155-670

Companies
Manufacturer
ALLERGAN
MARLOW INTERNATIONAL, PARKWAY
MARLOW, BUCKINGHAMSHIRE,
SL7 1YL
UNITED KINGDOM

Z. NATRELLE 410LF TruForm Soft Touch Silicone - Filled Breast Implant

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

N-ST-LF095-125
N-ST-LF100-150
N-ST-LF105-175
N-ST-LF110-205
N-ST-LF115-240
N-ST-LF120-270
N-ST-LF125-310
N-ST-LF130-350
N-ST-LF135-390
N-ST-LF140-440
N-ST-LF145-490
N-ST-LF150-540
N-ST-LF155-595

Companies
Manufacturer
ALLERGAN
MARLOW INTERNATIONAL, PARKWAY
MARLOW, BUCKINGHAMSHIRE,
SL7 1YL
UNITED KINGDOM

AA. NATRELLE 410LL TruForm Soft Touch Silicone - Filled Breast Implant

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

N-ST-LL105-135
N-ST-LL115-180
N-ST-LL120-210
N-ST-LL125-240
N-ST-LL135-300

Companies
Manufacturer
ALLERGAN
MARLOW INTERNATIONAL, PARKWAY
MARLOW, BUCKINGHAMSHIRE,
SL7 1YL
UNITED KINGDOM

AB. NATRELLE 410LM TruForm Soft Touch Silicone - Filled Breast Implant

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

N-ST-LM100-140
N-ST-LM110-190
N-ST-LM115-220
N-ST-LM120-250
N-ST-LM130-320

Companies
Manufacturer
ALLERGAN
MARLOW INTERNATIONAL, PARKWAY
MARLOW, BUCKINGHAMSHIRE,
SL7 1YL
UNITED KINGDOM

AC. NATRELLE 410MF TruForm Soft Touch Silicone - Filled Breast Implant

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

N-ST-MF095-140
N-ST-MF100-165
N-ST-MF105-195
N-ST-MF110-225
N-ST-MF115-255
N-ST-MF120-295
N-ST-MF125-335
N-ST-MF130-375
N-ST-MF135-420
N-ST-MF140-470
N-ST-MF145-525
N-ST-MF150-580
N-ST-MF155-640

Companies
Manufacturer
ALLERGAN
MARLOW INTERNATIONAL, PARKWAY
MARLOW, BUCKINGHAMSHIRE,
SL7 1YL
UNITED KINGDOM

AD. NATRELLE 410ML TruForm Soft Touch Silicone - Filled Breast Implant

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

N-ST-ML100-125
N-ST-ML110-170
N-ST-ML115-195
N-ST-ML120-220
N-ST-ML130-285

Companies
Manufacturer
ALLERGAN
MARLOW INTERNATIONAL, PARKWAY
MARLOW, BUCKINGHAMSHIRE,
SL7 1YL
UNITED KINGDOM

AE. NATRELLE 410MM TruForm Soft Touch Silicone - Filled Breast Implant

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

N-ST-MM095-135
N-ST-MM100-160
N-ST-MM105-185
N-ST-MM110-215
N-ST-MM115-245
N-ST-MM120-280
N-ST-MM125-320
N-ST-MM130-360
N-ST-MM135-400
N-ST-MM140-450

Companies
Manufacturer
ALLERGAN
MARLOW INTERNATIONAL, PARKWAY
MARLOW, BUCKINGHAMSHIRE,
SL7 1YL
UNITED KINGDOM