This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
NATALCARE ST-LX (2016-05-30)
- Starting date:
- May 30, 2016
- Posting date:
- June 16, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-58922
Affected products
- NATALCARE ST-LX
Reason
A possible defect in the rolling base of the NatalCare ST-LX incubator which could affect the safety of the product.
Affected products
A. NATALCARE ST-LX
Lot or serial number
- 5009
- 5490
- 5491
Model or catalog number
- NATAL CARE ST-LX
Companies
- Manufacturer
-
MEDIX I.C.S.A.
MARCOS SASTRE 1675
BUENOS AIRES, EL TALAR
B1618EWC
ARGENTINA