Health product recall

Nara Hospital Bassinet (2018-10-15)

Starting date:
October 15, 2018
Posting date:
October 26, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type I
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68108

Affected products

Nara Hospital Bassinet

Reason

It was identified [via customer complaint in May 2018] that the volume of air inside the mattress may expand in regions of high altitude which may result in an unintended gap between the plateau of the inflated patient surface and tub.

Affected products

Nara Hospital Bassinet

Lot or serial number

All lots.

Model or catalog number
  • 4402-000-002
Companies
Manufacturer
Stryker Medical
3800 E Centre Avenue
Portage
49002
UNITED STATES