Health product recall

Multix Fusion Digital (2018-08-08)

Starting date:
August 8, 2018
Posting date:
August 24, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67590

Affected products

Multix Fusion Digital

Reason

Due to a problem with the steel cables inside the lifting column of the overhead tube, the steel cables may rupture. The probability of occurrence depends on the degree of usage of the system and is rated as extremely rare. In case the cables break, the tube arm may fall down unexpectedly during movement in vertical direction.

Affected products

Multix Fusion Digital

Lot or serial number

40016

Model or catalog number

10893300

Companies
Manufacturer
Siemens Shanghai Medical Equipment Ltd.
278 Zhouzhu Road
Shanghai
201318
CHINA