Health product recall

MULTIDIAGNOST ELEVA IMAGE DETECTORS (2016-07-13)

Starting date:
July 13, 2016
Posting date:
July 29, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Healthcare Professionals, General Public, Hospitals
Identification number:
RA-59602

Affected products

  1. MULTIDIAGNOST ELEVA - IMAGE INTENSIFIER - TV IMAGE DETECTOR
  2. MULTIDIAGNOST ELEVA - FLAT DETECTOR - IMAGE DETECTOR

Reason

Philips has become aware that the two bolts that attach the clamping plate to the lateral shaft may break. This may cause the c-arm to float freely. Depending on the orientation of the free-floating C-arm and any other forces applied to it the C-arm could move laterally forward, slide out of its bearings and then possibly sag.

Affected products

A. MULTIDIAGNOST ELEVA - IMAGE INTENSIFIER - TV IMAGE DETECTOR

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

0708MD02

Companies
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684
NETHERLANDS

B. MULTIDIAGNOST ELEVA - FLAT DETECTOR - IMAGE DETECTOR

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

0708MD02

Companies
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684
NETHERLANDS