This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

MTS A/B/D Monoclonal and Reverse Grouping Card

Starting date:
November 5, 2012
Posting date:
December 10, 2012
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public (GP), Healthcare Professionals, Hospitals
Identification number:
RA-16279

Recalled Products

A. MTS A/B/D Monoclonal and Reverse Grouping Card

Reason

OCD has received complaints of an incomplete foil seal for specific lots of ID-Micro Typing System gel cards. Our investigation confirmed that some MTS gel cards exhibited a weaker or a loose foil seal resulting in the potential for drying of the first microtube. Using gel cards that exhibit drying in microtube(s) may cause false positive reactions leading to erroneous results.

Affected products

A. MTS A/B/D Monoclonal and Reverse Grouping Card

Lot or serial number

091312037-02

Model or catalog number

MTS080515

Companies
Manufacturer

Micro Typing Systems Inc.