Health product recall

MS-12 BLUE, CARDIOVIT CS-200 Office, Blue (2021-10-01)



Starting date:
October 1, 2021
Posting date:
October 15, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-76653


Last updated:

Affected Products

MS-12 BLUE, CARDIOVIT CS-200 Office, Blue

Reason

An error has been detected in the current software which does not indicate second/third degree AV-block interpretation in textual form. The software version below version 4.5.3 Implements a limited version of the resting ECG interpretation software.

Affected products

MS-12 BLUE, CARDIOVIT CS-200 Office, Blue

Lot or serial number

All lots.

Model or catalog number

3030.000100-3030.XXXXXX

Companies
Manufacturer

Schiller AG

Altgasse 68

Baar

6341

SWITZERLAND