MR SURGICAL SUITE II TABLE (2015-10-29)
- Starting date:
- October 29, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-55856
Affected Products
MR SURGICAL SUITE II TABLE
Reason
Potential safety issue with the table transfer release of the MR SURGICAL SUITE II MR System. The issue may cause an unexpected function loss of the patient transfer mechanism that allows the patient to be moved between the MR and surgical tables. This could delay care and handling of a patient during surgical or interventional procedures. GE has received one complaint of a transfer board not releasing from the MR table preventing the transfer board from moving to the surgical table. There have been no injuries reported as a result of this issue.
Affected products
MR SURGICAL SUITE II TABLE
Lot or serial number
416369MR3T
Model or catalog number
M0001SS
Companies
- Manufacturer
-
GE Medical Systems, LLC
3000 North Grandview Blvd.
Waukesha
53188
Wisconsin
UNITED STATES