Health product recall

MOSAIQ

Last updated

Summary

Product
MOSAIQ
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

MOSAIQ

Version 3.2.1.0 to 3.2.3.2

VERSION 3.2

Issue

When using Anatomical Position Monitoring (APM) during a Comprehensive Motion Management (CMM) unity treatment delivery, if the user performs a baseline shift (BLS) plan and an error occurs during the plan transfer, in some instances, the BLS plan may not load into mosaiq resulting in the original beams being delivered without the new BLS applied.

What you should do

Contact the manufacturer if you require additional information. 

  • Email: application.support.na@elekta.com
  • Tel: +1 855 6935358
  • https://www.elekta.com

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Elekta Solutions AB

Hagaplan 4, Stockholm, Sweden, 11368

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82038

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe