Health product recall

MOSAIQ (2018-12-03)

Starting date:
December 3, 2018
Posting date:
January 8, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68684

Affected products

MOSAIQ

Reason

If a setup field (CT, KV, MV OR MVCT) is created by copying a treatment field or another setup field, then the copied field will have the same isocenter as the parent and that isocenter may not be the correct isocenter for imaging when it is sent to the machine.

Affected products

MOSAIQ

Lot or serial number

Version 2.64 SP9 and higher

Model or catalog number
  • MOSAIQ
Companies
Manufacturer

Elekta Inc.

100 Mathilda Place

Sunnyvale

94086

California

UNITED STATES